VALIDATION OF IMPROVED HARDWARE AND SOFTWARE FOR EXPIRED GAS ANALYSIS INDIRECT CALORIMETRY

Show simple item record

dc.contributor.author Kim, Doyeon
dc.date.accessioned 2009-08-27T21:01:22Z
dc.date.available 2009-08-27T21:01:22Z
dc.date.issued 2009-08-27T21:01:22Z
dc.date.submitted July 2009
dc.identifier.uri http://hdl.handle.net/1928/9809
dc.description.abstract The purpose of this investigation was to validate a new system of breath-by-breath expired gas analysis to both an artificial working model of lung ventilation and gas exchange as well as to the Douglas bag technique. In addition, comparisons will be made between expired fractions, ventilation, and computations of VO2, VCO2, and RER between the new system and a commercial mixing chamber system (ParvoMedics) for repeated measurements at rest, steady state and non-steady state cycle ergometry exercise. Post acquisition processing involved custom developed software (LabVIEW), where time to gas equilibration within the mixing bag was determined, as well as differences in equilibrated gas fractions. All testing procedures were repeated 5 times for parametric statistical analyses. Gas concentration (%) results for the compliant 2 L mixing bag was the only method to yield data not significantly different between alveolar and measured. Alveolar % oxygen was significantly lower than mixing bag, mixing chamber, and ParvoMedics. The most responsive method was the mixing bag, with significantly lower % gas data for oxygen for breaths 2 to 5 compared to the mixing chamber and ParvoMedics. The ParvoMedics and mixing bag yielded similar results after breath 6, but data were significantly higher than for alveolar air. The slope data for breaths 0 to breaths 2 was significantly (p < 0.05) lower for the ParvoMedics system compared to the mixing bag and mixing chamber. The mean temporal distribution of 1 L ventilation maneuvers from the mixing bag turbine was 0.999 ± 0.142 L, with a range of 0.96 to 1.03 L. The mean ventilation (STPD) from the ParvoMedics (pneumotach) was significantly lower (p = 0.0027) than the mixing bag turbine. For VE (p = 0.097), VO2 (p = 0.786), and VCO2 (p = 0.178) were not significantly different in the main effect for method and the Intensity x Method interaction (VE: p = 0.721, VO2: p = 0.059, VCO2: p = 0.406). As expected, there was a significant difference for the intensity main effect (p < 0.0001). For FEO2 (p < 0.0001) and FECO2 (p < 0.0001) there were significant findings for the main effects of intensity. However, the Intensity x Method interaction showed no significant differences in FEO2 and FECO2. RER was significantly different in the main effect for method (p = 0.024), intensity (p = 0.0006), and Intensity x Method interaction (p = 0.005). The expired oxygen and carbon dioxide had significant main effects and interactions (p < 0.001). All mean differences between alveolar and mouth end tidal gas % values across 6 breaths were significant (p < 0.01). The mean individual computed dead space volumes were 2.5 ± 0.13 L. The results suggested that the new 2 L mixing bag is capable of accurately reproducing specific gas fractions from reference calibration gas. The new 2 L mixing bag allowed expired air to wash out through the bag. This system, in combination with including anatomical dead space (ADS) as a factor in the determinations, gives more accurate measurements and calculations than a traditional mixing chamber. Additionally, the new mixing bag method has unique aspects that are advantageous to the operation and validity of the system. Although the new system is not used in commercial systems of expired gas analysis indirect calorimetry (EGAIC), this system provides enhanced accuracy and validity. en_US
dc.language.iso en_US en_US
dc.subject EXPIRED GAS ANALYSIS en_US
dc.subject INDIRECT CALORIMETRY en_US
dc.subject expired fractions en_US
dc.subject Douglas bag en_US
dc.subject mixing chamber en_US
dc.subject Ve, VO2, VCO2, and RER en_US
dc.subject.lcsh Indirect calorimetry
dc.subject.lcsh Indirect calorimetry--Equipment and supplies--Evaluation
dc.title VALIDATION OF IMPROVED HARDWARE AND SOFTWARE FOR EXPIRED GAS ANALYSIS INDIRECT CALORIMETRY en_US
dc.type Dissertation en_US
dc.description.degree Physical Education, Sports and Exercise Science en_US
dc.description.level Doctoral en_US
dc.description.department University of New Mexico. Division of Physical Performance and Development en_US
dc.description.advisor Robert, Robergs
dc.description.committee-member Len, Kravitz
dc.description.committee-member Christine, Mermier
dc.description.committee-member Carole, Conn


Files in this item

Files Size Format View Description
Final.doc 6.281Mb Microsoft Word View/Open Main article
Microsoft Word - Final.pdf 864.0Kb PDF View/Open Main article(PDF)

This item appears in the following Collection(s)

Show simple item record

UNM Libraries

Search LoboVault


Advanced Search

Browse

My Account